Literature >
Basic Principles >
Basic Counting Systems -
Pulse Electronics
The nuclear electronics industry has standardized the signal
definitions, power supply voltages and physical dimensions of basic
nuclear instrumentation modules (NIM). The standardization provides
users with the ability to interchange modules, and the flexibility
to reconfigure or expand nuclear counting systems, as their counting
applications change or grow. Canberra is a leading supplier of Nuclear
Instrumentation Modules (NIM). Basic electronic principals,
components and configurations which are fundamental in solving
common nuclear applications are discussed below.
Preamplifiers and Amplifiers
Most detectors can be represented as a capacitor into which a
charge is deposited, (as shown in Figure 1.13). By applying detector
bias, an electric field is created which causes the charge carriers
to migrate and be collected. During the charge collection a small
current flows, and the voltage drop across the bias resistor is the
pulse voltage.

Figure 1.13: Basic Detector
and
Amplification
The preamplifier is isolated from the high voltage by a
capacitor. The rise time of the preamplifier’s output pulse is
related to the collection time of the charge, while the decay time
of the preamplifier’s output pulse is the RC time constant
characteristic of the preamplifier itself. Rise times range from a
few nanoseconds to a few microseconds, while decay times are usually
set at about 50 microseconds.
Charge-sensitive preamplifiers are commonly used for most solid
state detectors. In charge-sensitive preamplifiers, an output
voltage pulse is produced that is proportional to the input charge.
The output voltage is essentially independent of detector
capacitance, which is especially important in silicon charged
particle detection (i.e. PIPS detectors),
since the detector capacitance depends strongly upon the bias
voltage. However, noise is also affected by the capacitance.
To maximize performance, the preamplifier should be located at
the detector to reduce capacitance of the leads, which can degrade
the rise time as well as lower the effective signal size.
Additionally, the preamplifier also serves to provide a match
between the high impedance of the detector and the low impedance of
coaxial cables to the amplifier, which may be located at great
distances from the preamplifier.
The amplifier serves to shape the pulse as well as further
amplify it. The long delay time of the preamplifier pulse may not be
returned to zero voltage before another pulse occurs, so it is
important to shorten it and only preserve the detector information
in the pulse rise time. The RC clipping technique can be used in
which the pulse is differentiated to remove the slowly varying decay
time, and then integrated somewhat to reduce the noise. The unipolar
pulse that results is much shorter. The actual circuitry used is an
active filter for selected frequencies. A near-Gaussian pulse shape
is produced, yielding optimum signal-to-noise characteristics and
count rate performance.
A second differentiation produces a bipolar pulse. This bipolar
pulse has the advantage of nearly equal amounts of positive and
negative area, so that the net voltage is zero. When a bipolar pulse
passes from one stage of a circuit to another through a capacitor,
no charge is left on the capacitor between pulses. With a
unipolar pulse, the charge must leak off through associated
resistance, or be reset to zero with a baseline restorer.
Typical preamplifier pulses are shown in Figure 1.14.




Figure 1.14: Standard Pulse
Waveforms
The dashed line in the unipolar pulse indicates undershoot which
can occur when, at medium to high count rates, a substantial amount
of the amplifier’s output pulses begin to ride on the undershoot of
the previous pulse. If left uncorrected, the measured pulse
amplitudes for these pulses would be too low, and when added to
pulses whose amplitudes are correct, would lead to spectrum
broadening of peaks in acquired spectra. To compensate for this
effect, pole/zero cancellation quickly returns the pulse to the zero
baseline voltage.
The bipolar pulse has the further advantage over unipolar in that
the zero crossing point is nearly independent of time (relative to
the start of the pulse) for a wide range of amplitudes. This is very
useful in timing applications such as the ones discussed below.
However, the unipolar pulse has lower noise, and constant fraction
discriminators have been developed for timing with unipolar
pulses.
For further discussions on preamplifier and
amplifier
characteristics, please refer to each applicable module’s
subsection.
Pulse Height Analysis and Counting Techniques
Pulse Height Analysis may consist of a simple discriminator that
can be set above noise level and which produces a standard logic
pulse (see Figure 1.14) for use in a pulse counter or as gating
signal. However, most data consists of a range of pulse heights
of which only a small portion is of interest. One can employ either
of the following:
- Single Channel Analyzer and Counter
- Multichannel Analyzer
The single channel analyzer (SCA) has a lower and an upper level
discriminator, and produces an output logic pulse whenever an input
pulse falls between the discriminator levels. With this device, all
voltage pulses in a specific range can be selected and counted. If
additional voltage ranges are of interest, additional SCAs and
counters (i.e. scalers) can be added as required, but the expense
and complexity of multiple SCAs and counters usually make this
configuration impractical.
If a full voltage (i.e. energy) spectrum is desired, the SCA’s
discriminators can be set to a narrow range (i.e. window) and then
stepped through a range of voltages. If the counts are recorded and
plotted for each step, an energy spectrum will result. In a typical
example of the use of the Model 2030 SCA, the lower level
discriminator (LLD) and window can be manually or externally (for
instance, by a computer) incremented, and the counts recorded for
each step. However, the preferred method of collecting a full energy
spectrum is with a multichannel analyzer.
The multichannel analyzer (MCA), which can be considered as a
series of SCAs with incrementing narrow windows, basically consists
of an analog-to-digital converter (ADC), control logic, memory and
display. The multichannel analyzer collects pulses in all voltage
ranges at once and displays this information in real time, providing
a major improvement over SCA spectrum analysis.
Figure 1.15 illustrates a typical MCA block diagram. An input
energy pulse is checked to see if it is within the selected SCA
range, and then passed to the ADC. The ADC converts the pulse to a
number proportional to the energy of the event. This number is taken
to be the address of a memory location, and one count is added to
the contents of that memory location. After collecting data for some
period of time, the memory contains a list of numbers corresponding
to the number of pulses at each discrete voltage. The display reads
the memory contents vs. memory locations, which is equivalent to
number of pulses vs. energy.

Figure 1.15.
Multichannel Analyzer
Components
Counters and Ratemeters
Counters and ratemeters are used to record the number of logic
pulses, either on an individual basis as in a counter, or as an
average count rate as in a ratemeter. Counters and ratemeters are
built with very high count rate capabilities so that dead times are
minimized. Counters are usually used in combination with a timer
(either built-in, or external), so that the number of pulses per
unit of time are recorded. Ratemeters feature a built-in timer, so
that the count rate per unit of time is automatically displayed.
Whereas counters have a LED or LCD for the number of logic pulses,
ratemeters have a mechanical meter for real-time display of the
count rate. Typically, most counters are designed with 8-decade
count capacity and offer an optional external control/output
interface, while ratemeters are designed with linear or log count
rate scales, recorder outputs and optional alarm level
presets/outputs. Additional information may be found in the Counters and
Ratemeters Introduction section.
Miscellaneous Units
Various pulse processing functions have been incorporated into
NIM units, such as linear gates, pulse generators, pulse
stabilizers, etc. Many of these are described in the following
sections and in the introduction to each Nuclear
Instrumentation Modules (NIM) Section.
Simple Counting Systems
As related above, pulse height analysis can consist of a simple
single channel analyzer and counter, or a multichannel analyzer.
Generally, low resolution/high efficiency detectors (such as
proportional counters and NaI(Tl) detectors) are used in x ray or
low-energy gamma ray applications where only a few peaks occur. An
example of a simple NaI(Tl) detector-based counting system of this
type is illustrated in Figure 1.16.

Figure 1.16: NaI Detector and
Counter/Timer with Alarm Ratemeter
In this configuration, a Model 2015A Amplifier/SCA is used to
generate a logic pulse for every amplified (detector) pulse that
falls within the SCA’s "energy window". The logic pulse is then used
as an input to the Model 2071A Counter/Timer which provides the user
with a choice of either preset time or preset count operation. The
Model 2071A may be equipped with an optional Model 207X-03 EIA
Interface, which enables the Model 2071A to be read out to a
printer, or controlled and read out to a computer for data storage
or further analysis.
Alternatively, Model 1481LA Linear/Log Ratemeter is used as the
counter, with an alarm relay that will trigger if the count rate
exceeds a user preset value.
Although counters are still used in some applications, most of
today’s counting systems include a multichannel analyzer (MCA).
Besides being more cost effective than multiple SCA-based systems, a
MCA-based system can provide complete pulse height analysis such
that all nuclides, (i.e., even those not expected), can be easily
viewed and/or analyzed.
NaI(Tl)Detectors and Multichannel Analyzers
The need for a single-input Pulse Height Analysis with Sodium
Iodide detector is best served by a PC-card MCA, such as the
AccuSpec NaI/Plus (Figure 1.17).

Figure 1.17: AccuSpec/NaI Plus
MCA
Configuration
This single plug-in board includes a High Voltage Power supply,
Amplifier, and ADC in addition to its MCA functions, and thus, there
is no need for additional modules or a NIM Bin. Further technical
discussions concerning multichannel analyzers and multichannel
analysis systems (including spectroscopy software) may be found in
Multichannel
Analyzers and Advanced
Spectroscopy Software sections.
PIPS Detectors and Multichannel Analyzers
Alpha spectroscopy measurements of low-level samples require long
counting times. A large area PIPS detector, when configured with a
Canberra alpha spectrometer and multichannel analyzer, provides a
high resolution, low background, counting system that will satisfy a
multitude of alpha spectroscopy applications.
An example of a single chamber alpha spectroscopy system (that
can easily be upgraded) is illustrated in Figure 1.12. Note that the
Model 7401 Alpha Spectrometer is a complete, self-contained, 2-wide
NIM module that contains a vacuum chamber, vacuum gage, detector
bias supply, preamplifier/amplifier, SCA, counter/timer and pulser
for setup and test. Multiple Model 7401 Alpha Spectrometers can be
configured with a vacuum system that allows individual vacuum
chambers to be opened and loaded without affecting the vacuum or
data acquisition of the other spectrometers.
However, where numerous samples are counted simultaneously, it
may be more cost effective and user efficient to select a system
based on the Alpha Analyst (Figure 1.18).

Figure 1.18 Example Large Scale Alpha
Spectroscopy System
This turn-key system supports multiple detectors in a
comprehensive software environment featuring full computer control
of all vacuum elements and acquisition electronics. To learn more
about Canberra’s Alpha Analyst, click here.
Germanium Detectors and Multichannel Analyzers
A typical HPGe detector-based gamma spectroscopy system consists
of a HPGe detector, high voltage power supply, preamplifier (which
is usually sold as part of the detector), amplifier,
analog-to-digital converter and multichannel analyzer. Figure 1.19
illustrates a simple gamma spectroscopy system. This configuration
shows NIM electronics for the front end, allowing selection of the
optimal spectroscopy amplifier. Canberra offers traditional
‘manually-operated’ NIM modules, as well as a selection of
computer-controlled front ends.

Figure 1.19.
HPGe Detector and
MCA
For higher count rate applications, it is necessary to use an
additional circuit to reject pileup pulses that can distort the
spectrum. Pileup Rejection/Live Time Correction (PUR/LTC) inspects
both the leading edge and the trailing edge of the pulse and can
discriminate between two events separated by less than 0.5
microseconds. Since these pileup pulses are rejected, the ADC live
time must be lengthened to properly compensate for time the system
was unable to process pulses. Virtually all current Canberra ADCs
and MCAs (including AccuSpec, InSpector, and AIM) provide signal
paths for pulse Pileup Rejection/Live Time Correction as illustrated
in Figure 1.20.

Figure 1.20.
HPGe Detector with
PUR/LTC
LEGe and Si(Li) Detectors with Multichannel Analyzers
Low Energy
Germanium (LEGe) and Si(Li) detectors
require special circuitry to provide the long time constants
required in the amplifier to achieve maximum resolution, and to
properly handle the pulsed optical feedback signals of their
preamplifiers. Although several Canberra amplifiers are suitable,
the best resolution will be obtained with the Model 2025 AFT
Research Amplifier. Besides allowing the user to select a long
shaping time constant, the Model 2025 features an enhanced baseline
restorer which is ideal for pulsed optical feedback
preamplifiers.
In high count rate applications, the long time constants
contribute to Pulse Pileup. The Model 2025 contains a built-in Live
Time Corrector/Pileup Rejector to prevent these inaccuracies. A
typical example of a LEGe or Si(Li) based system is illustrated in
Figure 1.21. Note that this system also includes an optional Model
1786A Detector LN2 Monitor to prevent accidental damage
to the detector caused by running out of liquid nitrogen.

Figure 1.21.
LEGe or Si(Li) Detector
and MCA
Multiple Input Systems
Traditionally, Mixer/Routers, or Multiplexers, were employed to
allow several detectors to be counted in one MCA (with one ADC), as
shown in Figure 1.22. Advances in computer technology have
dramatically lowered the cost of MCAs and memory, so that, today, it
is frequently more effective to use multiple MCAs in place of a
Mixer/Router. The Mixer/Router configuration has a major drawback in
that a single ADC processes signals from all detectors, and thus the
count rate on the individual detectors must be relatively low to
avoid excessive pileup. Additionally, a Mixer/Router must allocate
the memory of the MCA among its inputs, which decreases the number
of channels available for an individual channel. Within these
constraints, Mixer/Routers can be quite efficient for applications
such as low-level environmental alpha spectroscopy, in which
multiple low-intensity inputs are collected in MCA memory segments
of 512 channels or less.

Figure 1.22.
Multiple Input
System
Low Level Gamma Ray Counting
Large volume HPGe detectors have become dominant over other
detector types for low level gamma ray spectroscopy because of their
inherently good resolution and linearity. It is only in the analysis
of single radionuclides that NaI(Tl) detectors can compare in
sensitivity with HPGe detectors. Since the majority of all gamma
spectroscopy applications require the analysis of more complex,
multi-radionuclide samples, the following discussion will be limited
to the application of HPGe detectors to low level counting.
The sensitivity of a HPGe spectrometer system depends on several
factors, including detector efficiency, detector resolution,
background radiation, sample constituency, sample geometry and
counting time. The following paragraphs discuss the role these
factors play in low level gamma ray counting.
- Efficiency
Generally, the sensitivity of a HPGe
system will be in direct proportion to the detector efficiency.
HPGe detectors are almost universally specified for efficiency
relative to a 3 in. NaI(Tl) at 25 cm detector-to-source distance
at 1.33 MeV, and from this benchmark one may roughly impute the
efficiency at lower energies. However, for the customer who is
counting weak samples with lower gamma energies, for instance
100-800 keV, the following subtle considerations to the detector
design are important to system performance:
- The detector should have an adequate diameter. This assures
that the efficiency at medium and low energies will be high
relative to the efficiency at 1.33 MeV, where it is bought and
paid for.
- The detector-to-end-cap distance should be minimal - five
millimeters or less. The inverse square law is real and will
affect sensitivity.
- The detector should be of closed end coaxial geometry, to
assure that the entire front face is active.
- Resolution
Generally, the
superior resolution of a HPGe detector is sufficient enough to
avoid the problem of peak convolution, (i.e., all peaks are
separate and distinct). The sensitivity of a system improves as
the resolution improves because higher resolution means that
spectral line widths are smaller, and fewer background counts are
therefore involved in calculating peak integrals.
Since the
sensitivity is inversely related to the square root of the
background, that is,

improvements in resolution will not improve sensitivity as
dramatically as increased efficiency.
- Background Radiation and Sample
Constituency
Interfering background in gamma spectra
originates either from within the sample being counted
(Compton-produced) or from the environment. If the sample being
analyzed has a high content of high-energy gamma emitting radioisotopes, the Compton-
produced background will easily outweigh the environmental
background. For extremely weak samples, the environmental
background becomes more significant. Obviously, massive
shielding will do little to improve system sensitivity for low
energy gamma rays in the presence of relatively intense higher
energy radiation. However, Compton-suppression can be very
effective in reducing this background.
- Sample Geometry
An often
overlooked aspect of HPGe detector sensitivity is the sample
geometry. For a given sample size (and the sample size should be
as a large as practicable for maximum sensitivity), the sample
should be distributed so as to minimize the distance between the
sample volume and the detector itself.
This rules out
analyzing "test tube" samples with non-well type detectors, or
"large area flat samples" with standard detectors. It does rule in
favor of using re-entrant or Marinelli-beaker-type sample
containers, which distribute part of the sample around the
circumference of the detector.
Germanium Detectors with Inert Shields
There are many different types of shield designs that are
available, and because of the difficulty in determining the
background contribution of the materials used in a given shield, it
is difficult to assign performance levels to various types of
shields. However, some criteria for shield designs have evolved over
the years, such as:
- The shield should not be designed to contain unnecessary
components like the Dewar. It will only contribute to increased
background if it is within the walls of the shield, as well as
unnecessarily increase the shield’s size, weight and cost.
- The detector should be readily installed and removable from
the shield.
Pity the person who has to move lead bricks (at
12 kg each) to disengage a HPGe detector. A HPGe detector and
shield system should have a liquid nitrogen transfer system to
avoid removing the detector for the weekly filling.
- Sample entry should be convenient to the operator.
- The shield should accommodate a variety of sample sizes and
configurations.
The HPGe detector should be located in the
center of the shield so as to minimize scatter from the walls. In
this position, the shield must accommodate the largest sample that
is anticipated. Also, sample placement should be accurately
repeatable and easily verified by the operator.
The shield design that has all these features and is moderately
priced is the Canberra Model 747 Lead Shield illustrated in Figures
1.23 and 1.24.

Figure 1.23 Detector located in center
of chamber without requirement for extended end-cap

Figure 1.24 Model 747 Lead
Shield
The performance of the shield using a Canberra HPGe detector is
given below:
| Shield Specs |
| Inside
Dimensions |
28 cm dia. x 40.5
cm high |
| Wall
Thickness |
10
cm |
| Material |
Low Background
Lead |
| HPGE Specs |
| Rel.
Efficiency |
12% |
| Resolution |
1.95
keV FWHM at 1.33 MeV 0.90 keV FWHM at 1.22
keV |
| Background
Count |
2.25
counts/second in the 50 keV-2.7 MeV
range |
Table 1.4 lists the sensitivities of several single
radioisotopes, assuming a counting time of 50 000 seconds, a 50%
error and a detector-to-point-source distance of 1 cm.
| Table 1.4 Radioisotope vs.
Sensitivity |
| Radionuclide |
Energy in keV |
Sensitivity in pC |
| 57Co |
122 |
2 |
| 139Ce |
165 |
3 |
| 137Cs |
662 |
6 |
| 60Co |
133 |
10 |
Low Background Cryostats
The design or configuration of the cryostat is another factor in
system performance. Some cryostat/shield designs do not prevent
streaming from the outside environment, nor do they provide
self-shielding from their own relatively hot components. Through an
improper choice of material types and/or thicknesses, the cryostat
may appreciably contribute to the background. Canberra has developed
sources for select, low-background, materials, and has invested in
the design and fabrication of low-background cryostats, as described
in the Introduction to the Cryostats and
Accessories Section.
HPGe Compton Suppression Spectrometer
When the ultimate in low level counting is required, a Compton
Suppression Spectrometer, in conjunction with an appropriate
low-background shield/cryostat design, is the answer. In this
configuration, the HPGe detector is surrounded by an active NaI(Tl)
or plastic scintillation guard detector (also known as an annulus
detector), with the electronics configured in an anticoincidence
counting mode. The Compton continuum, which is primarily caused by
gamma rays which sustain one or more inelastic collisions and escape
(i.e. scatter out of) the germanium detector material without
imparting their full energy, can lead to concealment of low activity
peaks. Since this is undesirable in low level counting applications,
a Compton Suppression Spectrometer can be used to gate (i.e. turn
off) data acquisition whenever one of the incompletely absorbed
photons escapes the germanium detector and is "seen" by the annulus
detector. When acquisition is complete, the resultant spectrum only
contains peaks attributed to gamma rays which have imparted their
full energy within the detector material.
It should be pointed out that some radioisotopes (those having
coincident gamma rays) such as 60Co, will not be analyzed
properly by the anticoincidence spectrum from a Compton Suppression
System. Therefore, two spectra are usually obtained from such a
spectrometer - one in the anticoincidence mode, and the other in the
normal (ungated) mode.
Figure 1.25 illustrates a typical example of
a Compton Suppression System.

Figure 1.25.
Compton Suppression
system
One type of annulus has six (6) 5.08 cm (2-inch) diameter
photomultiplier tubes (PMTs) on one end, and a 7.62 cm (3-inch)
diameter NaI(Tl) plug with one PMT (which is operated in parallel
with the other PMT) on the other end. A simpler type of annulus
detector uses a 15.24 cm (6-inch) diameter NaI(Tl) well detector on
a single PMT. In either configuration, the annulus must be large
enough to allow the insertion of the HPGe detector’s endcap along
with the sample.
While some endorse the use of a fairly complex Timing Chain to
derive the anti-Compton gate signal, Canberra has found that the
simplified circuit shown in Figure 1.25 yields equivalent results
(See the Compton Suppression Made Easy Application Note). The
"Incoming Count Rate" signals from the Spectroscopy Amplifiers are
checked for coincidence, and, if it exists, the 2040 Coincidence
Analyzer’s output is used as an anti-coincidence input to the ADC’s
Gate. When coincidence occurs, this gate "turns off" the delayed
unipolar signal from the Spectroscopy Amplifier. Typical Compton
Suppression Spectrometer results are illustrated in Figure 1.26.

Figure 1.26 Ge Spectra with
Compton
Suppression
It can be seen that the ‘figure of merit’ - the value of the
137Cs peak at 662 keV divided by the average contents of
the Compton continuum (the energy range 358-382 keV) - is on the
order of 1000:1.
High Count Rate Gamma Ray Systems
High count rate applications require special techniques to assure
good resolution and/or good throughput. In general, "high count
rate" is used to refer to incoming count rate, that is, the number
of events seen by the detector. The term "throughput rate" may be of
more interest to the researcher, being a measure of the rate at
which the system can accurately process these incoming counts.
In high count rate HPGe detector applications, problems such as
the loss of resolution, excessively long counting times, erroneous
peak to background ratios, inaccurate counting statistics or system
shutdown due to overload and saturation begin to appear. In some
experiments, the solution to these problems merely lies in reducing
the incoming count rate to the detector, or by employing electronics
which inhibit the processing of pulses through the electronics when
events are occurring so fast that they are overlapping (pulse
pileup). In this latter solution, system throughput will of course
be reduced, but parameters such as resolution will be enhanced.
Table 1.5 indicates the throughput limitations of the major
components of a spectroscopy chain. Note that the term "energy rate
limited" refers to the fact that the component’s performance is not
only affected by the incoming count rate, but by the relative energy
(amplitude) of the incoming counts as well. Each element in the
chain can be optimized for high count rate performance.

Table 1.5.
Major System
Components and their Throughput Limitations
The Detector
For the detector itself, the charge collection time is the
limiting factor, and this parameter is a function of the detector
geometry - when a photon interaction takes place, charge carriers in
the form of holes and electrons are produced, and the time taken for
these carriers to be swept to the P and N electrodes of the detector
is the time for full energy collection. In a Germanium detector,
this time is a function of detector size, as the charge carriers
travel about 0.1 mm/sec. As the charge collection time increases,
the Amplifier must take a longer time to process the signal and
develop its linear pulse, or else not all of the incident energy
will be reflected in that pulse ("ballistic deficit"). Thus, larger
detectors require longer amplifier time constants, or more
sophisticated peak shapes.
Some ways to address high count rate in the detector include
moving the detector farther away from the source, or collimating the
detector - which in both cases reduces the number of events seen by
the detector - or using a detector of less efficiency. The detector
in the latter case will ‘see’ fewer events, and furthermore will
have a lower charge collection time.
The Preamplifier
Most Germanium detectors in use today are equipped with
RC-feedback, charge sensitive preamplifiers. In the RC-feedback
preamplifier, a feedback resistor discharges the integrator,
typically in one or two milliseconds. If the incoming energy rate
(count rate X energy/count) produces a current that exceeds the
capability of the resistor to bleed it off, the output will increase
until, in the extreme, the preamplifier saturates and ceases to
operate. This limit occurs at approximately 200k MeV/s. The
saturated condition remains until the count rate is reduced. The
saturation limit is dependent on both energy and count rate and is
usually specified in terms of the "energy/rate limit". The
energy/rate limit can be increased by lowering the value of the
feedback resistor, but this in turn allows more noise to pass
through the preamplifier, resulting in a degradation in
resolution.
When a Coaxial Germanium detector is used in applications
requiring high throughput, the Model 2101 Transistor Reset
Preamplifier (TRP) is favored over traditional RC feedback
Preamplifiers. The higher cost of the TRP is justified by its much
higher energy rate capacity, an enhancement obtained by replacing
the Feedback Resistor of a typical RC feedback preamplifier with a
special reset circuit. This circuit monitors the dc level of the
preamplifier and discharges the feedback capacitor whenever the
output reaches a predetermined reset threshold. At moderate to high
count rates (i.e. above 20 000 cps), the absence of the feedback
resistor and its attendant noise and secondary time constant
contributions lead to:
- Lower preamplifier noise contributions.
- Inherently better resolution and reduced spectrum broadening
vs. count rate.
- Elimination of the need for pole-zero cancellation.
- Elimination of ‘lock-up’ due to saturation.
Figure 1.27 illustrates the throughout performance of the two
preamplifier styles.

Figure 1.27 Throughput vs. Count Rate:
Throughput Optimization
Although the Model 2101 TRP virtually never shuts down due to
saturation, its reset process and the amplifier overload which it
causes does induce intervals of dead time into the counting system.
The Model 2101 has been designed with a small Charge Gain (50
mV/MeV) and a wide Dynamic Range (4 V) to significantly reduce the
dead time due to resets in comparison to competitive units.
The Amplifier
The function of the spectroscopy amplifier is to filter the
signal for optimum signal to noise ratio and to provide gain, by
first differentiating the preamplifier output to provide a rapid
return to the baseline reference, and then integrating this signal
to capture the energy information. For the best resolution
performance, the integration stage produces a Gaussian waveform, and
the measure of the width of this pulse may be expressed in terms of
its time constant. If the time constant is too short, the amplifier
will allow more noise to pass, thus degrading resolution.
Additionally a short time constant can fail to give enough time
for the entire charge collection to take place in the detector
(ballistic deficit), which will also affect resolution. On the other
hand, a longer time constant lengthens the output pulse of the
amplifier (output pulse width is roughly equal to 7.3 x Time
Constant) and allows more pileup effects. Since pileup events - one
pulse riding on another - carry invalid information, they are
typically gated off from the amplifier output. Thus, decreasing the
time constant degrades resolution, and increasing it decreases
throughput. Figure 1.28 shows the effect of time constant on
throughput and resolution:


Figure 1.28
Resolution vs.
Throughput Tradeoffs for a Standard Spectroscopy
Amplifier
In practice, Coaxial Germanium detectors are usually not operated
below 2 ms Gaussian Shaping, and, more
typically, at
4 ms shaping to optimize
resolution.
A technique called Gated Integration (GI) can be used to allow
the use of short shaping time constants. The GI Amplifier integrates
the entire unipolar signal, providing a properly scaled Linear
Output signal with a relatively short time constant, while avoiding
the severe impact on resolution that would be seen with a Gaussian
pulse of the same width.
While the Gated Integrator, with shaping time constants on the
order of 0.25 ms, allows higher throughput,
it does introduce a certain amount of noise, and thus cannot achieve
the same resolution as a standard Gaussian output at, for example, a
shaping time of 4 ms. However, this slight
and constant degradation in resolution may be a desirable tradeoff
in terms of throughput, as evidenced by the plots of count rate and
resolution in Figure 1.29.


Figure 1.29 The Gated Integrator vs.
a Spectroscopy Amplifier
The ADC
Canberra offers two styles of ADC’s - Wilkinson and Fixed Dead
Time (FDT). The Wilkinson technique achieves better specifications
for differential and integral linearity, and hence resolution, but
its throughput is dependent not only on the count rate but by the
energy, or converted channel number, of the counts. For spectra with
conversion gain of 1024 channels or less, the Wilkinson ADC can
outperform most FDT ADC’s available today, with conversion times of
typically 5-10 ms. However, at higher
conversion gains, the FDT ADC’s will have a decidedly faster
conversion time, and thus higher throughput. The Canberra Model 8715
ADC, an FDT ADC with a conversion time of 800 ns, clearly offers the
best performance up to its conversion limit of 8K channels.
In performance tests, the Model 8715 FDT ADC was found to be so
fast that it actually adds NO dead time to spectroscopy
systems using traditional Gaussian or triangular shaping amplifiers,
and less than 800 nanoseconds to systems using Gated Integrator
amplifiers.
Summarizing, the combination of the RC Preamplifier, Model 2025
AFT Research Amplifier, and Model 8701 ADC provide optimum
resolution at incident counts rates less than 20 000 cps, but the
combination of Model 2101 TRP, Model 2024GI Amplifier, and Model
8715 FDT ADC will yield the highest throughput available - up to 70
000 cps, with only a minor tradeoff in resolution, as seen in Figure
1.29.
Pulse Pileup Rejection
As we saw earlier, pulse pileup occurs when a new pulse from the
preamplifier reaches the spectroscopy amplifier before the amplifier
has had a chance to finish properly processing the previous pulse.
In such cases, an Amplifier/ADC combination with Pile Up
Rejection/Live Time Correction (PUR/LTC) has the ability to (a)
inhibit the ADC from processing any invalid, composite, pulses, and
(b) turn off the Live Time Clock, which is measuring ‘real’ collect
time, until the next valid pulse is received at the Amplifier.
Typically, this clock is disabled during the time when the Amplifier
and ADC are processing a pulse, and thus records the time when the
system is ‘ready’ to receive an input pulse.
Digital Signal Processor
Historically, analog devices have been employed to perform the
function of the amplifier in Pulse Height Analysis systems. As noted
earlier, this function is not as much amplification as
it is processing or shaping. The
detector signal is processed, shaped, and filtered by the amplifier
and then digitized by the ADC at the end of the processing chain.
The selection of the output pulse shape and its associated time
constants are designed to maximize the signal-to-noise ratio for
optimized resolution and reduced sensitivity to ballistic deficit,
while providing the maximum throughput consistent with the
resolution requirements of the application. With advances in high
speed digital hardware, it is now possible to design and implement
digital pulse processors to perform the operations previously only
performed by an analog amplifier. With digital processing, filtering
functions can be employed that are not possible with conventional
analog signal processing.
In Canberra’s Digital Pulse Processing system, the
detector/preamplifier signal is digitized much earlier in the signal
processing chain. Subsequent processing, filtering, baseline
restoration, and pileup rejection are performed digitally, and the
results are transferred directly to the MCA histogram memory for
viewing and analysis. This system, introduced in 1997, has been
shown to provide resolution and throughput performance well in
excess of commercially available analog systems, as well as
increased precision and repeatability.
Loss Free Counting Applications
The correction of the Live Time Clock as described above,
effectively extending the counting time to account for those periods
when the system could not accept an input, is adequate for most
samples, in particular those for which the count rate is relatively
constant. However, for short Half-Lived samples, or samples whose
constituents change (as in a flow monitoring application), this
method will not be accurate. In addition, even if the "counts per
unit time" are accurate using the traditional method for dead time
correction, the "real" counting time will have been extended by an
amount equal to the dead time, which may in fact increase the actual
collection time to an undesirable length.
The principal goal of LFC is to insure that at the end of any
data acquisition interval, the electronics have accumulated all of
the events that occurred regardless of any dead time that may have
been present in the system. LFC is based on the concept that by
adding "n" counts per event to an MCA’s channel register, rather
than digitizing and storing a single count at a time, a "zero dead
time" energy spectrum can be accumulated that assures all counts are
included in the spectrum. Assuming that "n" is correctly derived,
("n" should equal "1" plus a "weighting factor" representing the
number of events that were lost since the last event was stored),
and the data is truly random in nature, the concept is statistically
valid. The factor "n" is derived on a continuous basis by examining
the state of the Amplifier and ADC every 200 ns. The proportion of
time during which the Amplifier and ADC are processing a pulse
provides a measure for the magnitude of the weighting factor "n",
which is updated every 20 ms. Loss free
counting requires that the MCA support "add-n" or multiple "add-one"
data transfer; consult the factory for details.
Unfortunately, counting statistics in a Loss Free Counting system
cannot be calculated from the corrected spectrum. One basic
assumption used by all peak fitting algorithms is that of Poisson
counting statistics. That is, the uncertainty of the counts is
proportional to the square root of the number of counts. While this
assumption is true for traditional "add-1" front-ends, it is not
true of the "add-n" Loss Free Counting front-end. This assumption is
especially poor as the weighting factor becomes large. To properly
quantify the uncertainty in each channel’s contents, the peak
fitting program must have access to both the corrected "add-n" and
the uncorrected "add-1" spectra. Therefore, Canberra also offers a
"Dual-LFC" hardware option for the Model 599 which allows the
collection of both of these spectra so that statistically correct
peak filling can occur. Note that the correction software for the
"Dual-LFC" system is only available for VMS-based Genie Systems.